商品信息

Description:
Quidel
Molecular Adenovirus Control Set
Features & Benefits:






The Lyra Adenovirus Assay is a real-time PCR
in vitro
diagnostic assay for the qualitative detection of human Adenovirus (HAdV) viral DNA which has been isolated from nasal and nasopharyngeal swabs obtained from individuals exhibiting signs and symptoms of acute respiratory infections. The test detects, but does not differentiate, HAdV species (A, B, C, D, E, and F) or
SERO
types.

During the test, a multiplex RT-PCR reaction is carried out under optimized conditions in a single tube generating amplicons for each of the target viruses present in the sample. Identification of adenovirus (HAdV) and the PRC occur by the use of target-specific primers and fluorescent-labeled probes which hybridize conserved regions in the genomes of adenovirus and the PRC.

The Lyra Adenovirus Assay has a convenient, one-step reagent set up performed at room temperature on the bench top. Moreover, optimized cycling conditions make results available in less than 75 minutes following extraction to aid in timely diagnosis and treatment. Lyra Adenovirus RT-PCR is available for use on the
Applied Biosystems
?
7500 Fast Dx Real-Time PCR Instrument.

The Lyra Adenovirus Assay, when used with other Lyra assays, provides the flexibility to customize respiratory panels with other Lyra Real-Time PCR Respiratory assays in order to adapt to the shifting demands caused by variable local or seasonal disease prevalence.










Feature
Benefit



One-step reagent set-up
Rehydration solution is simply added to the lyophilized Master Mix



Speed
Results in less than 75 minutes after extraction



Easy training
Simplified and uniform workflow with standard pipetting volumes



Refrigerated storage
No freezer needed. 2°C to 8°C storage



Long shelf life
2 years from date of manufacture



Room temperature set up
No ice or cooling block required



Flex
IBL
e 96-test format
Kit includes vials of 12x8 Master Mix, Rehydration Solution and liquid Process Controls



Multi-well multiplex panels
Standardized
Thermo
cycling conditions for the respiratory assays allow for multiple analytes to be analyzed on the same plate from the same extracted specimen, effectively providing a customizable panel of results












Product Specifications:




Product Spec Name
Product Spec Data



Sample type
Nasal swab, Nasopharyngeal swab



Time to results
Less than 75 minutes after extraction



Reagent storage conditions
2°C to 8°C



Controls storage conditions
2°C to 8°C



Sensitivity*
100%



Specificity*
95.5%



Positive Percent Agreement*
100%



Negative Percent Agreement*
98.7%



LOD
A/31: 8.00 x 10
-2
; B/3: 8.00 x 10
-2
; C/1: 8.00 x 10
-2
; D/19: 1.61 x 10
1
; E/4: 1.00 x 10
0
; F/41: 3.20 x 10
-2
(TCID
50
/mL)



Sample volume needed for extraction
180 ?L



Purified nucleic acid volume needed to testing
5 ?L



Shelf life
24 months








*Refer to Package Insert for additional performance claims.